Intek RheoVac 950 Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Ausrüstung Intek RheoVac 950 herunter. Intek RheoVac 950 User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 34
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
TABLE OF CONTENTS
SECTION 1 GENERAL INFORMATION ....................................... - 1 -
1.1 INTRODUCTION ................................................. - 1 -
1.2 PRINCIPLE OF OPERATION ....................................... - 2 -
1.3 TECHNICAL SPECIFICATIONS .................................... - 2 -
1.4 PRECAUTIONS .................................................. - 2 -
SECTION 2 INSTALLATION ................................................ - 4 -
2.1 INSTALLATION ................................................. - 4 -
2.2 RheoVac SYSTEM INSTALLATION/SITE SELECTION ................. - 4 -
2.2.1 Probe Site Selection .......................................... - 4 -
2.2.2 Electronics Unit Site Selection ................................. - 5 -
2.3 MOUNTING HARDWARE INSTALLATION .......................... - 6 -
2.4 PROBE INSTALLATION AND ORIENTATION ........................ - 6 -
2.5 ELECTRICAL CONNECTIONS ..................................... - 8 -
SECTION 3 OPERATION .................................................. - 14 -
3.1 GENERAL INFORMATION ....................................... - 14 -
3.2 SYSTEM START-UP ............................................. - 14 -
3.3 PORTABLE USB DATA STORAGE DEVICE/Warranty ................. - 14 -
3.4 DISPLAY ....................................................... - 15 -
3.5 COMMUNICATIONS ............................................ - 15 -
3.6 DATA PROCESSING ............................................. - 16 -
3.7 CUSTOM SOFTWARE ........................................... - 17 -
SECTION 4 MAINTENANCE .............................................. - 18 -
4.1 GENERAL MAINTENANCE ....................................... - 18 -
4.2 CALIBRATION ................................................. - 18 -
4.3 SPARE PARTS .................................................. - 18 -
4.4 TROUBLESHOOTING ............................................ - 18 -
4.5 HARDWARE AND SOFTWARE MODIFICATIONS AND UPGRADES . . . - 20 -
SECTION 5 CUSTOMER SERVICE ......................................... - 21 -
5.1 QUESTIONS ON EXISTING HARDWARE ........................... - 21 -
5.2 TROUBLESHOOTING ............................................ - 21 -
5.3 FACTORY AND FIELD SERVICE .................................. - 21 -
5.4 QUESTIONS ON NEW EQUIPMENT ............................... - 21 -
5.5 CONDENSER SERVICES ......................................... - 21 -
SECTION 6 CUSTOM INFORMATION ...................................... - 22 -
6.1 UNIT IDENTIFICATION .......................................... - 22 -
6.2 CONFIGURATION ............................................... - 22 -
6.3 SPECIAL INSTRUCTIONS ........................................ - 22 -
Appendix A - RheoVac Model 950 User Interface Software
Appendix B - RheoVac Model 950 Networking
Intek, Inc. 2007
Manual no. RV950 Rev. D
I:\OFFICE\WPMANUAL\RV950 R ev D.wpd
Seitenansicht 0
1 2 3 4 5 6 ... 33 34

Inhaltsverzeichnis

Seite 1 - TABLE OF CONTENTS

TABLE OF CONTENTSSECTION 1 GENERAL INFORMATION ... - 1 -1.1 INTRODUCTION ...

Seite 2

- 8 -Ñ Inspect the probe tips. Be sure wetted surfaces are clean before installing. If cleaning isneeded, use a damp cloth wetted with alcohol and wip

Seite 3 - the RheoVac

- 9 -Figure 5 Output Connections and Set-up

Seite 4 - ! 20 to 120°F (-29 to 49°C)

- 10 -Figure 6 Distribution Box4. Network Connection: The Ethernet connection at JP1 is an RJ-45 style jack. A 10 footEthernet Cat 5 crossover cable

Seite 5

- 11 -TABLE II - Optional 4-20m A C onfigurationChannelOutput Parameter1 Actual volume flow2 Total mass flow3 Water vapor flow4 Pressure5 Water vapor/

Seite 6 - SECTION 2 — INSTALLATION

- 12 -Figure 8 Optional Transmitter BoxFigure 9 Serial Communications Interface

Seite 7

- 13 -Figure 10 Probe Connector AssemblyE. Cable Connection from Probe to Distribution BoxThe RheoVac probes require special connectors to be install

Seite 8

- 14 -SECTION 3 — OPERATION3.1 GENERAL INFORMATIONThe RheoVac system is calibrated, compensated, and linearized for a wide range of flowing mediatem

Seite 9

- 15 -3.4 DISPLAYThe RheoVac 950 has a 2x20 alphanumeric scroll-through display that shows seven parameters for eachprobe. The display shows Air In-l

Seite 10 - S (transmitS )

- 16 -3.6 DATA PROCESSINGTABLE IV. Column Headings for Data Downloads and Archived Data*TimeStamp(Time ofday)ActualVolumeFlow(ACFM)Total MassFlow(lbs/

Seite 11

- 17 -3.7 CUSTOM SOFTWARECustom software may be developed by the user to receive and archive RheoVac data into a computersystem. The electronics has

Seite 12

WARRANTYIntek, Inc. warrants each RheoVac product to be free from defects inmaterial and workmanship under normal use and service, Intek'sobligat

Seite 13

- 18 -SECTION 4 — MAINTENANCE4.1 GENERAL MAINTENANCEPrecautions should be taken to insure proper performance of all sensors. Since the quantificati

Seite 14

- 19 -TABLE VI. Troubleshooting GuideMESSAGE CODEDescription/SymptomPROBABLE CAUSEACTIONMSG-0Communications notbeing received fromprobe(s)1. Improper

Seite 15

- 20 -TABLE VII. Troubleshooting Guide (software communication errors)OBSERVATION PROBABLE CAUSE ACTIONAppears the first time theapplication was exec

Seite 16 - SECTION 3 — OPERATION

- 21 -SECTION 5 — CUSTOMER SERVICEIntek's corporate philosophy is to help solve our customers’ difficult flow measurement problems.When you pu

Seite 17

- 22 -SECTION 6 — CUSTOM INFORMATION6.1 UNIT IDENTIFICATIONModel no.:Serial no.: Customer identification:6.2 CONFIGURATIONThe marked (X) items denot

Seite 18

- A1 -I:\OFFICE\WPMANUAL\RV950 APPENDIX A3.wpd January 29, 2007APPENDIX A - RheoVac® Instrument User Interface SoftwareA.1 INTRODUCTIONUser Interface

Seite 19

1File transfer can also be accomplished using Windows Explorer or Network Neighborhood2 Specific network set-up at Intek may be required.- A2 -I:\OFFI

Seite 20 - SECTION 4 — MAINTENANCE

- A3 -I:\OFFICE\WPMANUAL\RV950 APPENDIX A3.wpd January 29, 2007950" network ID). The user must select the network ID from the pull down menu fol

Seite 21

- A4 -I:\OFFICE\WPMANUAL\RV950 APPENDIX A3.wpd January 29, 2007Figure B - User Interface Software Data Monitor FeatureFigure C - User Interface Softwa

Seite 22

- A5 -I:\OFFICE\WPMANUAL\RV950 APPENDIX A3.wpd January 29, 2007This screen can be divided into four parts:1. The Chart (or charts)2. The X-Axis Contro

Seite 23

- 1 -Figure 1 RheoVac Model 950SECTION 1 — GENERAL INFORMATION1.1 INTRODUCTIONFor the first time, all necessary properties of the gases in the conden

Seite 24

- A6 -I:\OFFICE\WPMANUAL\RV950 APPENDIX A3.wpd January 29, 2007to the user’s hard disk is stored in the “C:\RHEOVAC\DATA\” directory. The data is sub

Seite 25 - APPENDIX A - RheoVac

I:\OFFICE\WPMANUAL\RV950 Appendix B2.wpd March 1, 2006-B1-APPENDIX B - RheoVac Model 950 NetworkingB.1 INTRODUCTIONUse these instructions when making

Seite 26

I:\OFFICE\WPMANUAL\RV950 Appendix B2.wpd March 1, 2006-B2-4. Networking with the RheoVac 950: If all of the items above are completed successfully, th

Seite 27

I:\OFFICE\WPMANUAL\RV950 Appendix B2.wpd March 1, 2006-B3-necessary. Note: The Windows installation disk may be needed as well as administratorsecuri

Seite 28

I:\OFFICE\WPMANUAL\RV950 Appendix B2.wpd March 1, 2006-B4-3. Check for Installed Network Components: Click the Start button and select Control Panel.R

Seite 29

- 2 -1.2 PRINCIPLE OF OPERATIONThe principal features of the RheoVac system are shown in Figure 1. At the heart of the RheoVacsystem is the Rheotherm

Seite 30 -  IMPORTANT 

- 3 -Use reasonable care in handling the probe the sensing components are delicate. Do notbend the probes, damage the tips, or obstruct the sensing

Seite 31

- 4 -Figure 2 RheoVac Probe Insertion RecommendationSECTION 2 — INSTALLATION2.1 INSTALLATIONThese instructions cover installation of RheoVac instrum

Seite 32

- 5 -# Check installation clearance. Each transducer probe is almost 3 feet (0.9m) long and the hottap assembly is approximately 13 inches (0.33m) lo

Seite 33

- 6 -Figure 3Probe Stop Clamp2.3 MOUNTING HARDWARE INSTALLATIONÎ Check hardware. Verify that the probe slides through the hot tap assembly.Ï Check ins

Seite 34

- 7 -Figure 4 Transducer Installation Detail

Kommentare zu diesen Handbüchern

Keine Kommentare