- 13 -
Figure 5. Typical Time Response Characteristics
CUSTOM INFORMA-TION.
3.4 OUTPUT CURVE
If the instrument is ordered with a calibrated
output signal, the output curve for the unit can
be found after SECTION 6. For an
uncalibrated unit with a 4-20mA option, a
typical curve is shown in Figure 6. Check
SECTION 6 for the corresponding full scale
value for this instrument. Flow can also be
found by the formula:
FLOW (%FS) = exp [ exp (0.5547* X + 0.26935 ) ] / 0.1758
where X is the flow output signal in mA.
Figure 6. Typical Output Curve (for units with 4-20mA output option)