WARRANTY
Intek, Inc. warrants each Rheotherm product to be free from defects
in material and workmanship under normal use and service, Intek's
obligation under this warranty being limited to making good any part
or parts thereof which shall, within one (1) year after delivery of such
product to the original purchaser, be returned to Intek with
transportation charges prepaid and which Intek's examination shall
disclose to its satisfaction to have been thus defective; this warranty
being expressly in lieu of all other warranties, express or implied and
all other obligation or liabilities on Intek's part. The purchaser will
assume all responsibility and expense for removal, decontamination
and reinstallation of equipment.
Rheotherm flow meters are manufactured under United States patent numbers 4,255,968; 4,942,763; 4,949,578; 5,485,754 and 5,752,411
Intek, Rheotherm, RheoVac, Rheovec, Rheomax and RheoSmart are registered trademarks of Intek, Inc.
Intek, Inc.
751 Intek Way
Westerville, Ohio 43082-9057
TEL: (614) 895-0301 • FAX: (614) 895-0319
website: www.intekflow.com
e-mail: sales@intekflow.com
Seitenansicht 0
1 2 3 4 5 6 ... 26 27

Inhaltsverzeichnis

Seite 1

WARRANTYIntek, Inc. warrants each Rheotherm product to be free from defectsin material and workmanship under normal use and service, Intek'soblig

Seite 2 - TABLE OF CONTENTS

-8-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdFigure 2 - Transmitter Terminal Connections2. Check the analog output configuration of the transmitter an

Seite 3

-9-I:\O FFICE\W PM A N U A L\M an200 rvb.w pd5. The status output, also located on field wiring terminal, JP7, is a digital 0-15Vdc output. Thisoutpu

Seite 4

-10-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdFigure 3. S ystem W iring D etail

Seite 5

-11-I:\O FFICE\W PM A N U A L\M an200 rvb.w pd2.4 SERIAL OUTPUTThe Model 200 has RS232/422 output receptacles used during factory calibration. There

Seite 6 - SECTION 2 ! INSTALLATION

-12-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdFigure 4 - Serial Communications Interface

Seite 7

-13-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdSECTION 3 ! OPERATION3.1 START UPTypically, the instruments come from the factory set up for the flow ra

Seite 8

-14-I:\O FFICE\W PM A N U A L\M an200 rvb.w pd1. Analog Outputs — The unit will have two 0-5Vdc, 0-10Vdc or 4-20 mA signals for flow andtemperature ou

Seite 9

-15-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdunplug the RJ-11 serial communication cable to avoid possible signalcontention while using the keypad.1.

Seite 10

-16-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdA message of ‘Flow Not Valid’ during calibration means either there is no flow, or flow is outof the mea

Seite 11

-17-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdSECTION 4 ! MAINTENANCE4.1 GENERAL MAINTENANCECertain precautions should be taken to insure proper perfo

Seite 12

TABLE OF CONTENTSSECTION 1 ! GENERAL INFORMATION ... -1-1.1 INTRODUCTION ...

Seite 13 - ! (transmit!)

-18-I:\O FFICE\W PM A N U A L\M an200 rvb.w pd1. The transducer continuity check is performed as follows (see Figure 6-A):A. Disconnect the transducer

Seite 14

-19-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdFigure 6 - Transducer / Electronics Test Configuration

Seite 15 - SECTION 3 ! OPERATION

-20-I:\O FFICE\W PM A N U A L\M an200 rvb.w pd4.5 TROUBLE SHOOTINGThe following tables provide easy-to-follow instructions to trouble shoot flow indic

Seite 16

-21-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdTABLE III. Trouble Shooting Guide - Instrument diagnosed problemsOBSERVATION PROBABLE CAUSE ACTION&apos

Seite 17

-22-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdRecord voltages in Table IV (last column) before contacting the factory. Be sure to use a high inputimpe

Seite 18

-23-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdSECTION 5 ! CUSTOMER SERVICEIntek's corporate philosophy is to solve our customer's difficul

Seite 19 - SECTION 4 ! MAINTENANCE

-24-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdSECTION 6 ! CUSTOM INFORMATION6.1 UNIT IDENTIFICATIONModel no.: Serial no.: Customer identification: 6

Seite 20

-25-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdTABLE OF ORIGINAL CALIBRATION DATAFOR FUNCTIONAL TESTUnit Serial Number - I II III I

Seite 21

-1-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdSECTION 1 ! GENERAL INFORMATION1.1 INTRODUCTIONThe Model 200 is a “smart” instrument having performance c

Seite 22

-2-I:\O FFICE\W PM A N U A L\M an200 rvb.w pd1.2 DESCRIPTION OF OPERATIONRheotherm flow meters are available with various nonintrusive and intrusive t

Seite 23

-3-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdProbes (NPT/2I, NPT/I, BF/2I, BF/I, etc.) — Take care not to bend the probes or damage thetips. Do not t

Seite 24

-4-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdSECTION 2 ! INSTALLATION2.1 TRANSDUCER!! CAUTION: All transducers have a directional arrow on the tag a

Seite 25

-5-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdSome TU transducers have an integrally mounted cable; do not pull on this cable, or attemptto remove the

Seite 26

-6-I:\O FFICE\W PM A N U A L\M an200 rvb.w pd2. Intrusive Probes —!! IMPORTANT: Recommended straight run for best accuracy is aminimum 20 diameter

Seite 27

-7-I:\O FFICE\W PM A N U A L\M an200 rvb.w pdThe transmitter housing should be installed keeping in mind the length and routing of the transducercable